The STUDIA UNIVERSITATIS BABEŞ-BOLYAI issue article summary

The summary of the selected article appears at the bottom of the page. In order to get back to the contents of the issue this article belongs to you have to access the link from the title. In order to see all the articles of the archive which have as author/co-author one of the authors mentioned below, you have to access the link from the author's name.

 
       
         
    STUDIA CHEMIA - Issue no. Special%20Issue%202 / 2009  
         
  Article:   COATED-WIRE TYPE ION SENSITIVE ELECTRODES BASED ON NANOSTRUCTURED ATO THIN FILM.

Authors:  .
 
       
         
  Abstract:  

The functional performances of some cation and anion sensitive self-made, coated-wire type ISEs (ion selective electrodes), where the conventional metallic conductor was replaced by nanostructured ATO (antimony doped tin dioxide) thin film are presented. Two different PVC based membranes, cation and anion sensitive, respectively, were coated on ATO films prepared by the sol-gel technique. By using AFM (Atomic Force Microscopy), EIS (Electrochemical Impedance Spectroscopy) and sheet resistance measurements the film nanostructured morphology and the good electrical conductivity, respectively, were established. The cation sensitive ISE shows the best functional characteristics towards K+: linear response range: 5´10-5-10-1ML-1, sensitivity: 56±1 DmV/DpC, detection limit: 2´10-5 ML-1. The anion sensitive ISE shows the best functional characteristics towards : linear response range: 3´10-5-10-1ML-1, sensitivity: 55±1DmV/DpC, detection limit: 1,8´10-5ML-1. The life time of the ISEs was 8 days.

 

Keywords: coated-wire type ISE, ATO, sol-gel.

 
         
     
         
         
      Back to previous page